(c) Enter the “warm start” sequence: Press S1, then
press [1] and [ENTER] on the keypad.
After you switch units and perform one of the above pro-
cedures, the Main Panel display will show “LOSS”, indi-
cating that nonvolatile memory is reset to the default
state and stored test data is lost.
Note
After you change operating units for a calibrated
system, you must recalibrate the load and strain
channels and reset all limits (Load, Extension,
and Strain).
Operating Units for Self Identified Load Cells
and Load Frames
The units that are used for load, extension and speed are
listed in Table 4-7.
Legend for Table 4-7:
LB = pounds N = Newtons
KLB = LB x 10 KN = N x 10
MLB = KLB x 10 MN =K N x 10
GF = grams force IN = inches
KGF = GF x 10 MIN = minutes
KKGF = KGF x 10 MM = millimeters
PARAMETER ENGLISH METRIC S.I.
*Load
LB
KLB
MLB
GF
KGF
KKGF
N
KN
MN
Extension
IN MM MM
Speed
IN/MIN MM/MIN MM/MIN
Table 4-7. Operating Units for Self-Identified Load Cells
Function of
Controls
Selecting Operating Units
4-33