(c) Use the [+/–] key to toggle between the enabled and
disabled state.
Set Pretension Level
To set the Pretension Level, do the following:
(a) Press [S1] [7] [5].
(b) After the [5] key has been pressed, read the current
Pretension Level on the Main Panel Display.
(c) To change the Pretension Level, enter the load value
for the Pretension Level on the numeric keypad on
the console. Press [ENTER] to complete the se-
quence.
Set Excess Tension Level
To set the Excess Tension Level, do the following:
(a) Press [S1] [7] [6].
(b) After the [6] key has been pressed, read the current
Excess Tension Level on the Main Panel Display.
(c) To change the Excess Tension Level, enter the load
value for the Excess Tension Level on the numeric
keypad on the console. Press [ENTER] to complete
the sequence.
Time Delays
As mentioned earlier, the grip status feedback to the con-
sole is from the solenoid, not the grip faces. A one sec-
ond time delay is used to allow the grips time to open
and close in the above sequences.
Air Kit Option M10-94400-1
6-50