EasyManuals Logo

Instron 4400 Series User Manual

Instron 4400 Series
271 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
limit value (maximum or minimum load or
strain) must also be negative. Press the +/- key
on the keypad to obtain the correct sign when
setting a limit.
2. If the sign (+ or -) of the output signal of the
load cell (or strain transducer) in use is nega-
tive (-) with increasing load (or strain), the
Series 4400 system considers the most nega-
tive value as the minimum load (or strain). For
example: If load limits of -300 lb and -100 lb are
specified for a test, set the minimum load
(LOAD MIN key) to -300 lb, and the maximum
load (LOAD MAX key) to -100 lb.
(c) Press ENTER on the keypad. The number stops flash-
ing.
(d) Press the CROSSHEAD ACTION key for the limit
just entered. Indicator LEDs (ACTION and
STATUS) for the limit will light.
(e) To set a crosshead action to occur at specimen break,
press the BREAK key and then a CROSSHEAD AC-
TION key.
EXAMPLE:
Crosshead to stop at a maximum load of 200 lb, and to
return to gauge length if specimen breaks.
Enter the key sequence:
[LOAD MIN] [2] [0] [0] [ENTER] [STOP] [BREAK]
[RETURN]
Whenever one of the limit condition LEDs, or the
BREAK LED, is lit, one of the four crosshead action
Electronic Limits M10-94400-1
5-32

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Instron 4400 Series and is the answer not in the manual?

Instron 4400 Series Specifications

General IconGeneral
BrandInstron
Model4400 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals