EasyManuals Logo

Instron 4400 Series User Manual

Instron 4400 Series
271 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #200 background imageLoading...
Page #200 background image
(a) When the “E” command displays, press [0] [EN-
TER]. The display shows “E 1. - -” if no error exists
in Section 1. If an error exists, a message displays
that relates to a failed test as shown in Table 8-2.
(b) Record any error message, then repeat pressing [EN-
TER] to display messages for any additional errors in
Section 1.
(c) After all error messages are shown for Section 1, the
program automatically steps to Section 2 when you
press [ENTER]. Continue the procedure, as in Steps
(a) and (b) above, for all sections through No. 8.
(d) After all error messages display for Section 8,
and you press [ENTER], the display shows “L”
for the Look command. Press [REJ] to return to
the “E” command.
(e) To escape from the Diagnostic Monitor, press “S1”
to start the Cyclic Self Test. When the Self Test
number displays, press [ENTER].
EXAMPLE NO. 2:
To display errors for a particular section:
(a) To select Section No. 3 error message list, for exam-
ple, when the “E” command displays, press [3] [EN-
TER]. The display shows “E 3. - -” if no error exists
on Section 3. If an error exists, a message displays
that relates to a failed test as shown in Table 8-2.
Error Messages M10-94400-1
8-6

Table of Contents