lows a detailed list of errors to be scanned which can di-
rectly indicate component failures.
Note
The Diagnostic Monitor is for use by qualified
maintenance personnel, but limited use is de-
scribed in Chapter 9 for the readout of error
messages.
If the Diagnostic Monitor mode is entered accidentally
(by pressing S1 once when the Self Test Routine
number is displayed, or by pressing S1 then [2] on the
keypad during normal operation), the word HELP will
appear on the Main Panel Display. This mode can be ex-
ited to resume normal operation, if there are no failures,
by either pressing REJ (Reject) or by the following pro-
cedure:
Press S1 to enter the Cyclic Self Test, wait until the Self
Test result number is displayed, then press ENTER on
the keypad.
Resident Test Program Overview M10-94400-1
4-22