Maintain the Handpiece
9-4 CUSA EXcel Ultrasonic Surgical System Service Manual
Clean the Handpiece
Items needed: Soft cloths (one moist, one dry), swab or pipe cleaner, mild
cleansing solution, and rinse water.
Procedure:
1. Using a soft cloth, manually clean the handpiece, handpiece cable, and
standard nosecone with a mild cleansing solution (neutral detergent) or blood
dissolving detergent according to hospital policy.
2. Using a swab or pipe cleaner, manually clean the internal threads and suction
port of the connecting body.
3. Using a soft cloth, manually clean the internal threads and face of the
connecting body. Refer to Figure 9-1.
Figure 9-1.
Cleaning the handpiece with a soft
cloth
4.
Rinse the handpiece cable, handpiece housing, connecting body, and nosecone
thoroughly with water.
5. Using a soft cloth, dry the handpiece cable, handpiece housing, connecting
body, and standard nosecone.
Notice
Product damage will result if you do not follow these notices when cleaning the
handpiece:
Do not immerse the handpiece cable electrical connector in liquid.
Do not use ultrasonic or automated cleaners.
Do not use chlorinated substances such as bleach solution.
Do not clean the handpiece with abrasives such as Ajax, Comet, or steel wool.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals