Suction Pump
Calibration
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual 12-3
Suction Pump
1. Remove the left cover.
2. Remove the left end panel from the suction pump enclosure.
3. On the control panel, set the unit to Run mode.
4. Apply the digital tachometer to the suction pump motor shaft.
5. Adjust the Max. speed pot on the suction pump controller card so that the
motor operates at 3200 ± 20 rpm.
6. On the control panel, set the unit in Standby mode.
7. With the digital tachometer, verify that motor speed is greater than 1200 rpm.
8. Apply Torque Seal compound (Integra part number xxxx232 301 176xxxx) to
the Max. speed pot.
9. On the suction pump enclosure, replace the left end panel.
10. On the console, replace the left cover.
Motor for the Cooling Water Pump
Set Up
Before calibrating the motor for the cooling water pump, prepare the cooling
water system:
1. Fill and attach the cooling water reservoir.
2. Attach a handpiece to the console.
3. Turn on the AC Main Switch (rear panel) and the System Power Switch (front
panel).
When you turn on the System Power Switch, the CUSA EXcel system begins
to circulate cooling water through the handpiece. This takes about one minute
to complete. While the system is performing this task, the amber Wait
indicator flashes and the Cooling Water indicator (upper right on the control
panel) illuminates.
When the system turns the Wait and Cooling Water indicators Off, it turns On
the suction pump, the Test button LED, and the Prime button LED. The
cooling water system is ready for calibration.
4. Prime the system: Press the Prime button (blue box on a black button at the
top of the Irrigation display).
Watch for irrigation fluid to drip from the handpiece tip. Priming takes about
one minute.
After priming, the system automatically turns the Prime button LED Off, then
goes to the Standby status.
Procedure