Maintain the Handpiece
Maintaining the
CUSA EXcel System
CUSA EXcel Ultrasonic Surgical System Service Manual 9-5
Lubricate the Handpiece Connector O-Rings
Keeping the o-rings lubricated prevents them from drying out and cracking,
which may cause handpiece cooling problems.
Items needed: A high viscosity lubricant such as silicone grease.
Items to be lubricated: The four o-rings on the handpiece connector water
prongs.
Replace the Handpiece Connector O-Rings
Regularly replacing the o-rings prevents excessive wear, which may result in
handpiece cooling problems.
Items needed: O-Ring maintenance kit (Part Number 260-650-000), which
includes o-rings, an insertion tool, and instructions.
Items to be replaced: The four o-rings on the handpiece connector water prongs.
Refer to Figure 9-2.
Figure 9-2.
Replacing the handpiece connector
o-rings
Recalibrate the Handpiece
Handpieces for the CUSA EXcel system do not need recalibration unless a
problem develops. If this happens, contact Integra to arrange returning the
handpiece for service. (Refer to Return Equipment for Service in Section 13,
Repair Policy.)
O-ring
Insertion toolHandpiece connector

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals