Suction Pinch Valve
11-40 CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Install the Suction Pinch Valve
1. Replace the pinch valve by pushing it into place.
Note: Ensure that the electrical leads are free to move.
2. Replace the pinch valve electrical leads on the cable restraints.
3. Replace the two screws (M4 Phillips) that hold the pinch valve onto the pinch
valve collar.
4. Connect the suction pinch valve to the ultrasonics PCB (J18).
5. Replace the console rear and left covers.
Install the Irrigation Pump Bezel
1. Align the 4 ball-end retainers that hold the irrigation pump bezel in place.
Attach the pump bezel by pushing it straight toward the console.
2. Close the irrigation pump latch by rotating the latch lever clockwise until it
reaches the 3 o’clock position.
3. Plug the power cord connector into the wall receptacle.
4. Turn on the AC Main Switch, located on the rear panel.