Console Body Side Panel
Console Components
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual 4-11
Control Panel
Refer to pages 4-4 through 4-7.
Rotation Lock Handle
This handle releases the control panel, allowing it to swivel and lock into one of
four positions:
• Straight forward
• 45° right
• 45° left
• 90° left.
Housing for Irrigation Pump Head and Suction Pinch Valve
Refer to pages 4-2 and 4-3.
I.V. Pole
Refer to pages 4-2 and 4-3.
I.V. Pole Release Button
This button releases the I.V. pole, allowing you to raise it or lower it.
Arm and Arm Housing
Refer to pages 4-2 and 4-3.
Arm Release Tab
This tab releases a lock that holds the arm in three vertical positions.
Cooling Water Reservoir
This container holds cooling water to be continuously circulated through the
handpiece. The reservoir holds 1000 cc of distilled water.
A clear panel, which runs vertically up the side, allows you to see the water level
inside the reservoir.
The reservoir also includes fittings that snap into receptacles inside the Arm
Housing on the console side panel. The fittings and receptacles automatically
connect the reservoir to the cooling water system.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals