SECTION
8
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual 8-1
8Checkout Procedures
In this section:
• Inspections
Footswitch
Power cord
• Power on verification
• Handpiece test (optional).
Inspections
When you have unpacked the console, and at regular intervals afterward, inspect
the footswitch and the power cord.
Footswitch
1. Disconnect the footswitch from the console (rear panel).
2. Inspect the connector for damage or corrosion.
3. Inspect the footswitch for damage.
4. Inspect the cable for cuts, tears, or excessive compression.
5. Clean the footswitch and cable with a damp cloth and mild detergent.
6. Allow the footswitch to dry.
7. Connect the footswitch to the console (rear panel).