Tools/Equipment Needed for Calibrations
12-2 CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Tools/Equipment Needed for Calibrations
The calibrations require one or more of the following tools:
• Tachometer
• Irrigation bottle with fluid and administration set
• Digital multimeter, Fluke 87 or equivalent
• Vacuum gauge with T fitting and hose, 0–30 in. Hg
• Ashcroft 5 in. Dial-Grade 3A 0.25% or equivalent
• Suction canister
• Contamination guard
• Slotted blade screwdriver
• Phillips screwdriver.
DC Bias Power Supply
1. Leave all power off.
2. Remove the rear panel.
3. Put DVM leads between pins 1 and 6 of the output jack on the DC bias power
supply. You will find it on the AC input assembly (facing the rear of the
console, you see it on the left side of the assembly).
4. Turn on the AC Main Switch (rear panel) and the System Power Switch (front
panel).
5. Adjust the potentiometer to 4.7 Vdc (-0 + 0.05).
6. Turn off the System Power Switch (front panel) and the AC Main Switch (rear
panel).
7. Apply a suitable potentiometer locking compound to the adjustment screw of
the potentiometer.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals