Console Covers
Replacement Procedures
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual 11-5
Re-install the Left Cover
1. Position the bottom of the cover on the left side of the console and tilted
slightly outward.
2. Push the top of the cover inward toward the console. When the cover is in
place, push the left side latch toward the rear of the console.
3. Replace the left side lock screw (M5 Phillips), located under the console top
near the left rear corner.
Re-install the Right Cover
1. Position the bottom of the cover on the right side of the console and tilted
slightly outward. Be sure that the cooling water line service loop fits into the
console side.
2. Push the top of the cover inward toward the console. When the cover is in
place, push the right side latch toward the rear of the console.
3. Replace the right side lock screw (M5 Phillips), located under the console top
near the right rear corner.
4. Replace the cooling water reservoir.
Re-install the Rear Cover
1. Position the bottom of the cover on the rear of the console and tilted slightly
outward. Allow it to slide downward and to hook into the console.
2. Hold the top of the cover against the console and replace the two screws (M5
Phillips).
3. Reconnect the footswitch connector to the receptacle on the rear panel.