Console Body Rear Panel
Console Components
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual 4-13
Speaker
The speaker sounds an audible tone when an alert condition exists and when
vibration is active.
Volum e Control
This potentiometer allows you to adjust the active vibration audible tone. You
cannot adjust the alert tone.
Footswitch Connector Receptacle
This receptacle allows you to connect the footswitch to the console.
AC Main Switch
This switch controls the flow of AC power to the CUSA EXcel system. It does not
activate the system; instead, it allows power to flow from the wall receptacle to
the System Power Switch.
Once the AC Main Switch is On ( | ), you can leave it On even when you unplug
the system from the wall receptacle.
Fuses
These protect the system from electrical overloads. Refer to Section 11,
Replacement Procedures for instructions on replacing the fuses.
Footswitch
The footswitch has two pedals in a base, and allows you to activate two functions:
Ultrasonics/vibration: Right pedal (orange), when pressed, activates vibration at
the tip. When you release the pedal, vibration stops.
Fast Flush: Left pedal (blue), when pressed, increases the irrigation rate to
25 to 30 cc/min. It also closes the Suction Pinch Valve, stopping suction. When
you release the pedal, irrigation returns to the setpoint value and suction
resumes.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals