EasyManuals Logo

Integra Cusa Excel Service Manual

Integra Cusa Excel
213 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #173 background imageLoading...
Page #173 background image
Suction Pinch Valve
11-38 CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Suction Pinch Valve
Accessing the suction pinch valve requires removing the irrigation pump bezel.
The following procedures describe how to remove and re-install the irrigation
pump bezel and the suction pinch valve.
Tools Needed
You will need a Phillips head screwdriver.
Remove the Irrigation Pump Bezel
1. Turn off the AC Main Switch, located on the rear panel.
2. Unplug the power cord connector from the wall receptacle.
3. Open the irrigation pump latch halfway so that the latch lever stands at the 12
o’clock position.
4. Four ball-end retainers hold the irrigation pump bezel in place. Remove the
pump bezel by pulling it straight forward. Refer to Figure 11-27.
Remove the Suction Pinch Valve
1. Remove the console rear and left covers.
2. Disconnect the suction pinch valve from the ultrasonics PCB (J18).
3. Remove the two screws (M4 Phillips) that hold the pinch valve onto the pinch
valve collar.
4. Remove the pinch valve electrical leads from the cable restraints.
5. Remove the pinch valve by pulling it forward. Refer to Figure 11-28.
Note: Ensure that the electrical leads are free to move.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals