System 24 Volt Power Supply
11-6 CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
System 24 Volt Power Supply
Replacing the system 24 volt power supply involves removing the left side cover,
disconnecting wiring harnesses, and removing the power supply chassis.
Installing a new power supply reverses the process.
Tools Needed
• Phillips screwdriver
• 7 mm nut driver.
Remove the System 24 Volt Power Supply
1. Turn off the AC Main Switch, located on the rear panel.
2. Unplug the power cord connector from the wall receptacle.
3. Remove the left side cover.
4. Disconnect the wiring harness (refer to Figure 11-2).
5. Remove two hex nuts (M4), two lock washers, and two flat washers to release
the power supply chassis from the console.
6. Slide the power supply and its chassis out the left side of the console.
7. Remove the power supply and its chassis from the console.
Install the System 24 Volt Power Supply
1. Replace the power supply in the console.
2. Slide the power supply and its chassis into the console.
3. Replace the two hex nuts (M4), two lock washers, and two flat washers to
fasten the power supply to the console.
4. Connect the wiring harness. Refer to Figure 11-2.
5. Replace the left side cover.
6. Plug the power cord connector into a wall receptacle.
7. Turn on the AC Main Switch, located on the rear panel.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals