User Troubleshooting
Troubleshooting
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual 10-11
Little or no vibration or fragmentation
(Amplitude LEDs may illuminate well
below setpoint)—continued
Tip loose due to improper assembly 1. Press the Status button to place the
system in Standby mode.
2. Press the Test button to test the
handpiece and tip.
3. If the Vibration alert activates, do one
of the following:
If tip assembled in a nonsterile area:
a. Remove the handpiece from the
sterile field.
b. Use the tip torquing base and
wrench to tighten the tip.
c. Test the tip.
d. Sterilize the handpiece.
If sterile tip assembled to handpiece on
the sterile field:
a. Use the sterilizable torque base
and a disposable sterile wrench to
tighten the tip. (Refer to the
Supplement, Assembling or
Changing Tips on the Sterile
Field, in the CUSA EXcel
Ultrasonic Surgical Aspirator
System Users Guide.)
b. Test the tip.
Condition Causes Recommended Action

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals