SERVICE MANUAL
Invacare
®
- Kite
24
PROBLEM OTHER
SYMPTOMS
POSSIBLE
CAUSE
SOLUTION Documentation
Mobility
device
judders in
drive mode
None
Batteries
defective
(unstable
voltage)
Replace batteries See chapter 8.4.
Drive motor(s)
defective
Replace motor(s) See chapter 8.10.1.
Replace carbon
brushes
See chapter 8.10.4.
Batteries
not being
charged
None Batteries
defective
Replace batteries See chapter 8.4.
LEDs blinking
on charging
unit
Charging unit
defective
Replace charging
unit
See charging unit
operating manual
Mobility
device runs
too slowly
None Remote
defective
Replace remote See chapter 8.2.5.
Batteries
defective
Replace batteries See chapter 8.4.
Electrical
actuator
does not
respond
Remote
displays
flashing "E",
status diode
on
light/actuator
module does
not extinguish
even if the
remote is shut
down or
disconnected
Lighting/actuat
or module
faulty
Replace
lighting/actuator
module
See chapter 8.1.5.
None Cable
disconnected
or damaged
Ensure correct
cable
connection,
replace cable if
necessary
See chapter 8.6.
Electrical
actuator faulty
Check actuator See chapter 7.3.
Remote faulty Replacing the
remote
See chapter 8.2.5.