EasyManua.ls Logo

Invacare Kite - Actuator Checking Procedure

Invacare Kite
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Invacare
®
- Kite
SERVICE MANUAL
29
7.3 Checking an actuator
Requirements:
digital multimeter with resistance measurement
Switch the electronics OFF on the remote.
Remove the rear cover, as described in chapter 8.1.1.
Take exact note of the positions of all cables and the sockets that they are connected to. Mark
the connectors and sockets or take a photograph with a digital camera.
Disconnect the motor plug (1) of the motor to be tested from the actuator module.
Check the electrical resistance at the
plug (1) of the actuator.
The plug can have a different shape than
show in the illustration.
If the resistance is close to infinity, the motor
is probably burnt out.
If the resistance is below 1 , the motor has
a short-circuit.
The motor must be replaced in both cases.

Table of Contents

Other manuals for Invacare Kite