EasyManua.ls Logo

ITech IT8000 Series

ITech IT8000 Series
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MEASure Subsystem
9 MEASure Subsystem
Measure commands measure the input voltage or current. They trigger the ac-
quisition of new data before returning the reading. Measurements are per-
formed by digitizing the instantaneous input voltage or current for a specified
measurement time, storing the results in a buffer, and calculating the value for
the specified measurement type.
MEASure[:SCALar]:CURRent[:DC]?
Initiates, triggers, and returns the averaged current measurement. Values re-
turned in amperes.
Group
MEASure
Syntax
MEASure[:SCALar]:CURRent[:DC]?
Arguments
None
Default Value
None
Returns
<NRf+>
Example
MEASure[:SCALar]:CURRent[:DC]?
Also see
None
Copyright © Itech Electronic Co., Ltd.
103

Table of Contents

Other manuals for ITech IT8000 Series

Related product manuals