TRIGger Subsystem
Group
TRIGger
Syntax
TRIGger:ACQuire:VOLTage:HYSTeresis:LOW <NRf+>
Arguments
<NRf+>
MIN|MAX|DEF|<value>
Setting range: MIN to MAX
Default Value
MIN
Example
TRIGger:ACQuire:VOLTage:HYSTeresis:LOW 10
Query Syntax
TRIGger:ACQuire:VOLTage:HYSTeresis:LOW? [MIN|MAX|DEF]
Returns
<NRf+>
MIN|MAX|DEF|<value>
TRIGger:ACQuire:CURRent:SLOPe <CPD>
This command is used to set the current trigger edge of the Meter.
POSitive: Rising edge
NEGative: Falling edge
EITHer: Both rising and falling edges
Before executing this command, you need to set the trigger source of the
Meter function to current.