T able 6.1
U ltraviolet/visible region bandw idth (U V ) nm N ear-infrared bandw idth (N IR ) nm
0.1 0.4
0.2 0.8
0.5 2
1 4
2 (L2) 8
5 (L5) 20
10 (L10) 40
[W avelength] S ets the m easurem ent w avelength. Input range varies according to the
m odel.
V -530: 190.0 to 1100.0 nm
V -550/560: 190.0 to 900.0 nm
V -570: 190.0 to 2500.0 nm
[E nd Tim e] M easurem ent tim e interval in seconds.
[D ata P itch] D ata collecting w avelength interval. S electable range varies according to
the m odel.
V -530: 0.1, 0.2, 0.5, 1, 2 nm
V -550/560/570: 0.025, 0.05, 0.1, 0.2, 0.5, 1, 2 nm
[S am ple N o.] S ets w hich sam ple should be m easured first. S ubsequent sam ples are
m easured in order from the first sam ple num ber, increasing in increm ents
of one.
[D isplay] S ets the upper and low er lim its of the vertical axis range displayed on the
screen. If [A uto] is selected, the full-scale axis is set to about 1.2 tim es the
m axim um w idth of the displayed spectrum , based on the m easurem ent
result.
<O K > E nds param eter setting, and closes the dialog box. C lick this button to
transfer the set param eters to the spectrophotom eter.
<C ancel> S tops param eter setting w ithout changing the previous settings, and
closes the dialog box.
<S ave...> Saves param eters in the param eter library on the hard disk (see S ection
5.1.2.1 P aram eters] dialog box).
<O pen...> P reviously saved param eters in the param eter library can be selected (see
S ection 5.1.2.1 [P aram eters] dialog box).
6.1.2.2 [Data File] dialog box
C lick [D ata F ile] in the [P aram eters] dialog box to display the [D ata F ile] dialog box (see S ection 5.1.2.2
[D ata F ile] dialog box).
6.1.3 [M ove W avelength...]
M oves the spectrophotom eter w avelength to a set w avelength (see S ection 5.1.4 [M ove W avelength...]).
6.1.4 [A uto Zero]
S ets the absorbance (or transm ittance) at the current w avelength to zero (or 100% T for transm ittance).
6.1.5 [E xit]
E xits the [T im e C ourse M easurem ent] program and returns to [S pectra M anager].