Chapter 3 Ethernet Testing
OAM service and link layer testing
98
HST-3000 Ethernet Testing User’s Guide
4 Go to the S-OAM AIS configuration menu, and then spec-
ifying the following:
Specify Domain Select one of the follow-
ing:
– If you are testing per
IEEE 802.1ag, select
Yes.
– If you are testing per
ITU-T Rec. Y.1731,
select No.
MD ID
(Specify Domain must be
yes)
If you indicated that you
want to specify a domain
ID, enter the ID using up
to 22 characters.
The instrument uses the
ID that you specify to indi-
cate whether CCMs are
detected with different
IDs.
MA ID Specify the Maintenance
Association ID, using up
to 25 characters.
The instrument uses the
ID that you specify to indi-
cate whether CCMs are
detected with different
IDs.
Setting Parameters
AIS State If you want to test AIS,
select On; otherwise,
select Off, and proceed to
step 5.
NOTE: When you turn this
On, AIS will be transmitted
continuously.
Setting Parameters