Chapter 7 Automated RFC 2544 Testing
Running the Expert RFC 2544 test
214
HST-3000 Ethernet Testing User’s Guide
Running the Expert RFC 2544 test
Running the Expert RFC 2544 test involves the following:
Launching a single stream terminate application.
Specifying the RFC 2544 Mode on the Test Mode configu-
ration menu.
Specifying the RFC 2544 settings, such as the load
format (bit rate or percentage) and length type (frame or
packet lengths).
Enabling each of the tests that you want to run during the
course of the RFC 2544 test.
Asymmetric testing for PPPoE and IPv6 traffic is not
supported in this release.
When running the Expert RFC 2544 test, the Start Traffic
action is turned off, because the test automatically transmits
traffic when you start the test.
Running the
test
Before running the Expert RFC 2544 test, verify that no other
applications are running.
To run the Expert RFC 2544 test
1 If you haven’t already done so, launch a single stream
terminate application for the circuit you are testing (see
“Launching an application” on page 28).
2 Initialize the link (see “Initializing the link for Ethernet
testing” on page 47).
3 If you are using the JDSU Discovery feature to find your
instrument’s test partner (and configure key settings on
your instrument), press the Home button to return to the
Summary Results screen, press the Action soft key, then
select Discover Units. For details on this feature, refer to
“Discovering another JDSU test instrument” on page 166.