Chapter 7 Automated RFC 2544 Testing
Running the Expert RFC 2544 test
HST-3000 Ethernet Testing Users Guide
215
4 Use the right arrow key to display the Test Mode configu-
ration menu, and then do the following:
a Specify the standard test mode settings (see
“Speci-
fying test mode and network visibility settings” on
page 30).
b In RFC 2544 Mode, select Symmetric,
Asym Upstream, Asym Downstream, or
Asym Combined. If you intend to run the System
Recovery test, you must select Symmetric mode.
5 Use the right arrow key to display the RFC 2544 Settings
configuration menu, and then specify the following:
Setting Parameter
Load Format Bit Rate
Percentage
AutoSave Enable
Disable
Customer If you want the customer name to
appear in report output, enter a
name.
Technician If you want the technician name to
appear in report output, enter a
name.
Location If you want the location where you
launched the test to appear in
report output, enter it.
Comments If you want to include comments in
the report output, enter them.
NOTE: The following address settings are specified on the
IP Init menu if you run the test using layer 3 or layer 4 traffic.
They appear on the Network Visibility menu when you run
the test using layer 2 traffic.