Features and Functions 3
Keysight 33210A User’s Guide 131
10 Define the remaining waveform points.
Using the Time and Voltage softkeys, define the remaining waveform points
using the values shown in the table below.
The time value of the last point that can be defined in the waveform must
be less than the specified cycle period.
The waveform editor automatically connects the last waveform point to the
voltage level of first point to create a continuous waveform.
To insert additional points after the current waveform point, press the
Insert Point softkey. The new point is inserted midway between the current
point and the next defined point.
To remove the current waveform point, press the Remove Point softkey.
The remaining points are joined using the interpolation method currently
selected. You cannot remove Point #1 since the waveform must have a
defined initial value.
Point Time Value Voltage Value
10 s0 V
2 2 ms 3 V
3 4 ms 1 V
4 7 ms 0 V