Remote Interface Reference 4
Keysight 33210A User’s Guide 225
SWEep:SPACing {LINear|LOGarithmic}
SWEep:SPACing?
Select linear or logarithmic spacing for the sweep. The default is Linear. The
:SPAC? query returns “LIN” or “LOG”.
For a
linear
sweep, the function generator varies the output frequency
in a
linear fashion during the sweep.
–For a logarithmic sweep, the function generator varies the output frequency in
a logarithmic fashion during the sweep.
SWEep:TIME {<seconds>|MINimum|MAXimum}
SWEep:TIME? [MINimum|MAXimum]
Set the number of seconds required to sweep from the start frequency to the stop
frequency. Select from 1 ms to 500 seconds. The default is 1 second. MIN = 1 ms.
MAX = 500 seconds. The :TIME? query returns the sweep time in seconds.
The number of discrete frequency points in the sweep is automatically
calculated by the function generator and is based on the sweep time that you
select.
SWEep:STATe {OFF|ON}
SWEep:STATe?
Disable or enable the sweep mode. To avoid multiple waveform changes, you can
enable the sweep mode after you have set up the other sweep parameters.
The default is OFF. The :STAT? query returns “0” (OFF) or “1” (ON).
The function generator will not allow the sweep mode to be enabled at the
same time that burst or any modulation mode is enabled. When you enable
sweep, the burst or modulation mode is turned off.