4 Remote Interface Reference
254 Keysight 33210A User’s Guide
MEMory:STATe:RECall:AUTO {OFF|ON}
MEMory:STATe:RECall:AUTO?
Disable or enable the automatic recall of the power-down state from
storage
location “0” when power is turned on. Select “ON” to automatically
recall the
power-down state when power is turned on. Select “OFF” (default) to issue a reset
(*RST command) when power is turned on (state “0” is not automatically
recalled). The :AUTO? query returns “0” (OFF) or “1” (ON).
MEMory:STATe:VALid? {0|1|2|3|4}
Query the specified storage location to determine if a valid state is currently
stored in that location. You can use this command before sending the *RCL
command to determine if a state has been previously stored in this location.
Returns “0” if no state has been stored or if it has
been deleted. Returns “1” if a
valid state is stored in the specified location.
MEMory:STATe:CATalog?
Returns the names assigned to storage locations 0 - 4. If no user-defined names
have been assigned, the query returns the factory defaults:
AUTO_RECALL”,“STATE_1”,“STATE_2”,“STATE_3”,“STATE_4”,
This command returns the factory default name for location 0 (“AUTO_RECALL”)
unless a new name has been defined. Although you can define a name for
location 0, the name is overwritten when the power is cycled and a new power
down state is stored in this location.
MEMory:NSTates?
Query the total number of memory locations available for state storage. Always
returns “5” (memory location “0” is included).