EasyManua.ls Logo

Keysight Technologies 33210A - Page 300

Keysight Technologies 33210A
367 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Error Messages
300 Keysight 33210A User’s Guide
-221
Settings conflict;
trigger output connector used by burst gate
If you have selected the gated burst mode (BURS:MODE GAT command) with burst
enabled, the “trigger out” signal cannot be enabled (OUTP:TRIG ON command).
The rear-panel Trig connector cannot be used for both operations at the same
time.
-221
Settings conflict;
trigger output connector used by trigger external
When the
external
trigger source is selected (
TRIG:SOUR EXT
command
),
the function generator automatically disables the “trigger out” signal.
The rear-
panel
Trig
connector cannot be used for both operations at the same
time
.
-221
Settings conflict;
frequency reduced for user function
For arbitrary waveforms, the output frequency is limited to 3 MHz. When you
change to the arbitrary waveform function (APPL:USER or FUNC:USER
command)
from a function that allows a higher frequency, the function generator will
automatically adjust the frequency to 3 MHz.
-221
Settings conflict;
frequency changed for pulse function
For pulse waveforms, the output frequency is limited to 5 MHz. When you change
to the pulse function (APPL:PULS or FUNC:PULS command) from a function that
allows a higher frequency, the function generator will automatically adjust the
frequency to 5 MHz.