Error Messages 5
Keysight 33210A User’s Guide 305
-221
Settings conflict;
not able to sweep arb waveforms, sweep turned off
The function generator cannot generate a sweep using the arbitrary waveform
function. The sweep mode has been turned off.
-221
Settings conflict;
pulse wid th decreased due to period
For a pulse waveform, the function generator will automatically adjust the
waveform parameters in the following order as needed to generate a valid pulse:
(1) edge time, (2) pulse width or duty cycle, and then (3) period.
In this
case, the function generator has decreased the pulse width to accommodate
the specified period (the edge time is already at its minimum setting).
-221
Settings conflict;
pulse duty cycle decreased due to period
For a pulse waveform, the function generator will automatically adjust the
waveform parameters in the following order as needed to generate a valid pulse:
(1) edge time, (2) pulse width or duty cycle, and then (3) period.
In this
case, the function generator has decreased the pulse duty cycle to
accommodate
the specified period (the edge time is already at its minimum
setting).
-221
Settings conflict;
edge time decreased due to period
For a pulse waveform, the function generator will automatically adjust the
waveform parameters in the following order as needed to generate a valid pulse:
(1) edge time, (2) pulse width or duty cycle, and then (3) period.
In this case, the function generator has decreased the edge time to accommodate
the specified period and preserve the pulse width setting.