Remote Programming Reference
81150A and 81160A User’s Guide 345
Command
:DATA[1|2]:MOD:COPY
Long
:DATA[1|2]:MODulation:COPY
Parameters
<destination arb name> [,VOLATILE]
Parameter Suffix
Description
Copies the waveform from VOLATILE memory to the specified name in non-
volatile memory. The source for the copy is always VOLATILE. You cannot
copy from any other source and you cannot copy to VOLATILE. The arb
name may contain up to 12 characters. The first character must be a letter
(A-Z), but the remaining characters can be numbers (0-9) or the underscore
character (“ _ ”). Blank spaces are not allowed. If you specify a name with
more than 12 characters, a Program mnemonic too long” error is
generated.
Example
:DATA:MOD:COPY ARB_1,VOLATILE