Remote Programming Reference
81150A and 81160A User’s Guide 411
Command
:PWM[1|2]:INT:FUNC[?]
Long
[:SOURCE]:PWM[1|2]:INTernal:FUNCtion[:SHAPe][?]
Parameters
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|NOISe|USER}
Parameter Suffix
Description
Select the shape of the modulating waveform. Used only when the Internal
modulation source is selected (PWM[1|2]:SOUR INT command). (The
carrier must be a pulse waveform for PWM.) The default is SIN. The :
PWM[1|2]:INT:FUNC? query returns “SIN, “SQU”, “RAMP,
“NRAM”, “TRI,NOIS or USER”.
Select “SQU” for a square waveform with a 50% duty cycle.
Select “RAMP” for a ramp waveform with 100% symmetry.
Select “TRI” for a ramp waveform with 50% symmetry.
Select “NRAM” (negative ramp) for a ramp waveform with 0%
symmetry.
If you select an arbitrary waveform as the modulating waveshape (USER”),
the waveform is automatically limited to 16K (16384) points. Extra waveform
points are removed using decimation.
Example
:PWM:INT:FUNC NOIS