Remote Programming Reference
81150A and 81160A User’s Guide 497
Command
:SWE[1|2]:STAT[?]
Long
[:SOURce]:SWEep[1|2]:STATe[?]
Parameters
{ON|OFF}
Parameter Suffix
Description
Disable or enable the sweep mode. To avoid multiple waveform changes,
you can enable the sweep mode after you have set up the other sweep
parameters. The default is OFF. The :SWE[1|2]:STAT? query returns
“0” (OFF) or “1” (ON).
The instrument will not allow the sweep mode to be enabled at the same
time that burst or any modulation mode is enabled. When you enable
sweep, the burst or modulation mode is turned off.
Example
:SWE2:STAT OFF