Service Manual
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
82 Revision 2
Risk of material damage in the event of light failure!
-identical battery packs can lead to light failure or shorten the service life of the light.
Always replace the battery packs as a set (2 items).
Danger of injury due to short circuit! Explosion hazard!
Be sure to use only genuine Gebrüder Martin battery packs!
Non-compliance will void any warranty claims against Gebrüder Martin.
Returning used battery packs!
Batteries whose capacity has dropped below 80% of their rated capacity (for more information, see
marLED® battery, Instructions for Use) must be marked “To be checked by supplier” before returning
them.
Open contacts must be protected with masking tape.
Use suitable transport packaging. If undamaged, the original packaging can be used for this purpose.
• Pack tightly so that the batteries cannot move around in transit.
• Remove the sensoGrip.
• Open the light housing, see section 9.8 “Opening the Light Housing”, page 79.
This includes disconnecting the battery cables, see section 9.5 “Disconnecting the Battery Cable
(marLED® E battery)”, page 77.