3-7
Smart532
System
Description
3.1.8. System Specifications
3.1.8.1.
Outputs
3.1.8.2.
Operation and
Control
Treatment Laser
Type: Frequency-doubled Nd:gdVO4,
diode- pumped solid-state
Wavelength: Green: 532 nm ±2 nm
Nominal Laser Peak Power
1
: 50 to 2,500 mW
Exposure Duration: Single pulse
2
/Continuous Waveform: 5.0 to 3,000 ms
SmartPulse Waveform: 0.05 to 1.0 ms
Exposure Interval: Continuous Waveform: 10 to 1,500 ms
SmartPulse Waveform: 0.5 to 10 ms
Aiming Laser
Type: Red diode laser
Wavelength: 635 nm ±5 nm
Intensity: 7 settings from minimal to 1 mW max.
Operational mode: Continuous
Control
Microprocessor based, touch screen, high resolution, multicolor GUI.
Presets storing and display capability.
Laser Emission Control
Footswitch
Laser Emission Indicators
Blue LED
Audible indicator
1
At Repeat mode the peak power is limited by duty cycle limitation (refer to the Power
Setting section).
2
Single Pulse is a sub set of Continues one, with all the supporting work ranges, but
where the interval is zero.