EasyManua.ls Logo

LUMENIS smart532 - 3.1.8. System Specifications; 3.1.8.1. Outputs; 3.1.8.2. Operation and Control

LUMENIS smart532
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UM-1151140EN, Rev. C, May 2016
3-7
Smart532
™
System
Description
3.1.8. System Specifications
3.1.8.1.
Outputs
3.1.8.2.
Operation and
Control
Treatment Laser
●
Type: Frequency-doubled Nd:gdVO4,
diode- pumped solid-state
●
Wavelength: Green: 532 nm ±2 nm
●
Nominal Laser Peak Power
1
: 50 to 2,500 mW
●
Exposure Duration: Single pulse
2
/Continuous Waveform: 5.0 to 3,000 ms
SmartPulse Waveform: 0.05 to 1.0 ms
●
Exposure Interval: Continuous Waveform: 10 to 1,500 ms
SmartPulse Waveform: 0.5 to 10 ms
Aiming Laser
●
Type: Red diode laser
●
Wavelength: 635 nm ±5 nm
●
Intensity: 7 settings from minimal to 1 mW max.
●
Operational mode: Continuous
Control
●
Microprocessor based, touch screen, high resolution, multicolor GUI.
●
Presets storing and display capability.
Laser Emission Control
●
Footswitch
Laser Emission Indicators
●
Blue LED
●
Audible indicator
1
At Repeat mode the peak power is limited by duty cycle limitation (refer to the Power
Setting section).
2
Single Pulse is a sub set of Continues one, with all the supporting work ranges, but
where the interval is zero.

Table of Contents