UM-1151140EN, Rev. C, May 2016
5.4. Pre-Operative
Instructions
1.
Verify that the laser system is properly connected as instructed in this
manual.
2.
Verify that the delivery system is properly connected as instructed in
the delivery system operator manual. If using a remote interlock,
verify that it is also properly connected.
3.
Post the Laser in Use warning sign outside the treatment room door.
4.
Plug the laser system’s main power cable into the wall socket.
5.
Ensure that all persons in the treatment room are wearing the
appropriate laser safety eyewear. See Laser Safety Eyewear in Chapter
2 of this manual for detailed laser safety eyewear information.
6.
Turn on the laser system.
7.
Perform the aiming beam integrity check as instructed later in this
chapter.
● Lasers generate a highly concentrated beam of light that may
cause injury if improperly used. To protect the patient and the
operating personnel, the entire laser and the appropriate
delivery system operator manuals, including all Safety and
Regulatory sections, should be carefully read and
comprehended before operation.
● Do not use the delivery system if the aiming beam is set to
high intensity and is weak or not visible; the fiber optic cable
may be damaged. A damaged fiber optic cable may cause
accidental laser exposure or injury to the treatment room
personnel or patient, and/or fire in the treatment room.
● Before using the laser, always verify that the displays on the
laser console are properly working.