UM-1151140EN, Rev. C, May 2016
Maintenance
and
Troubleshooting
6.2. Professional
Maintenance
This section covers tests, calibrations and maintenance that require internal
access to the Smart532 console and special skills.
These procedures demand specific knowledge, training and
use of tools not available to repair personnel outside of
Lumenis. Since performing these procedures may expose the
user to potential electrical and laser energy hazards, Lumenis
requires that these procedures only be performed by trained
service personnel.
6.2.1. Power Meter
Calibration
6.2.2. Internal Power
Meter Test
Regulatory agencies require that manufacturers of US FDA CDRH Class
II and IV, and European IEC 60825 Class 2 and 4 medical lasers supply
their customers with power calibration instructions.
Power meter check and calibration must be performed by an engineer or
technician qualified to work with laser equipment. Questions regarding
this procedure should be referred to your Lumenis representative.
Disclaimer warning:
Calibration is a service procedure to be performed only by Lumenis-
certified service engineers. Adjustment by anyone other than a trained
Lumenis service engineer voids any existing manufacturer's warranty on
the instrument. It is company policy not to distribute service tools outside
of the Lumenis Service Organization. Possession of service instructions or
tools does not authorize repair or modification of a Lumenis instrument by
uncertified personnel.
The Smart532 laser system incorporates an internal power meter which is
used for display and control of lasing energy. The power meter test
compares the internal power meter reading to the reading from an external
power meter.
Optical components must be clean before the power meter check is
performed.
● All personnel in the immediate area must wear eye protection
rated specifically for the 532 nm laser.
● An out-of-calibration power meter will cause power delivery to
tissue to be different than that displayed.