UM-1151140EN, Rev. C, May 2016
System
Installation
and
Initial
Setup
4.3.3. Environmental
Requirements
Air Quality:
The system should operate in a non-corrosive atmosphere. Corrosive
materials such as acids can damage electrical wiring, electronic
components and the surfaces of optical components.
Air-borne dust particles should be kept to a minimum. Dust particles
absorb light and heat up. Hot particles located on the optical lenses can
damage them. Metallic dust is destructive to electrical equipment.
Temperature:
To ensure that the system performs optimally, it is recommended to
maintain ambient room temperature between 10-30°C (50-86°F) and
relative humidity of 10-75% non-condensing. When the system is used
intensively it will emit heat. Therefore, it is recommended that the
treatment room be air-conditioned.