UM-1151140EN, Rev. C, May 2016
Maintenance
and
Troubleshooting
Table 6-1: Troubleshooting Guide: Error Messages with Corrective Actions (continued)
# Error Message Text Corrective Action
24
Footswitch Error
Smart footswitch failure; check cable connection
and retry, replace or repair footswitch.
25
Footswitch Error
Power Ease footswitch failure; check cable
connection and retry, replace or repair footswitch.
30
Power Converge Error (SubP)
System could not generate requested power.
Restart system and try again; If error persists, shut
down the system and contact Lumenis Service.
32.
Door Open (Interlock)
Close the interlocked door or connect the interlock
plug. Access Ready mode and continue normal
operation.
33
Emergency Button Pressed
Pull the laser emergency stop button out and restart
the system.
34
On/Off Button Pressed
Release the button and resume normal operation.
35
Invalid Password
Enter the correct login password and resume
normal operation.
39
Temperature Range Error (System)
The ambient temperature in the treatment room is
outside of the system's specs. Adjust the room
temperature and restart the system.
42
Front Port: Accessory Identification
Error
Incompatible delivery device connected.
43
Rear Port: Accessory Identification
Error
Incompatible delivery device connected.
44
No Delivery Device Connected
Connect a compatible delivery device and resume
normal operation.
45
Two LIOs connected
Disconnect one of the LIOs and resume normal
operation.
46
Two Laser Links connected
Disconnect one of the laser links and resume
normal operation.
47
Two Probes connected
Disconnect one of the probes and resume normal
operation.