Root NIBPT Chapter 5: NIBP Measurement
www.masimo.com 101 Masimo
Set Mode: Schedule
Schedule* interval measurement mode will take blood pressure measurements once every
defined interval for the defined duration. A patient profile can have up to five (5) schedules.
The schedules will run consecutively, when one schedule ends the next one will begin
automatically.
*Schedules are configured by authorized personnel.
To measure blood pressure in Schedule mode:
1. Ensure that the correct patient profile is selected before measurement.
WARNING: Only use Root in Neonatal mode with a neonatal blood pressure cuff to
measure blood pressure on neonates.
2. To change the patient profile, press the Main Menu icon , then select Profiles
.
3. Press the Profile Name to select the desired patient profile.
4. Properly place the blood pressure cuff on the patient. See Cuff Selection and
Placement on page 93.
5. To enable Schedule mode, press the Main Menu icon , then select NIBP
Settings .
6. On the Intervals screen, change Set Mode to Schedule. The set mode can also be
changed using the action menu.
WARNING: Too frequent blood pressure measurements can cause injury to the
patient due to blood flow interference.