Root NIBPT Chapter 15: Specifications
www.masimo.com 141 Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Emissions
Guidance and Manufacturer's Declarations - Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an
environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
ME Equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A
For hospital environment only.
Not intended for use in a domestic environment.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies