Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 150 Masimo
Power-On Self Test
To conduct a Power-On Self Test
1. Connect Root to AC power, and verify that the AC Power Indicator is illuminated.
2. Power On Root. Within five (5) seconds, all available indicators will illuminate, the
device will emit a tone, and the Masimo logo will display.
NIBP Module Calibration Test
Note: This section is provided as a reference and intended for authorized service personnel
only.
Pass Criteria
International standards for automated NIBP devices require that the maximum static
pressure accuracy shall be ± 3mmHg or 2% or the reading, whichever is greater. This is a
stringent requirement and all test equipment must be in excellent working order to properly
perform this test. It is important to verify the calibration before changing it. Historical data
has shown that the transducers rarely need to be re-calibrated although we still suggest that
the calibration be verified annually.
Equipment
Calibrated Manometer*
Pneumatic "T" Adapters
500mL bottle or regular-sized cuff wrapped around solid object
Hand Bulb
Connection Tubing
*Verify the manometer has been calibrated within the last 12 months. Calibration accuracy
shall be within ±0.02% to ±0.25% FS (Full Scale) and a measurement uncertainty of ±1 LSD
(Least Significant Digit).

Table of Contents

Other manuals for Masimo Root

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Root and is the answer not in the manual?

Masimo Root Specifications

General IconGeneral
CategoryMedical Equipment
Display10.1 inch touchscreen
ConnectivityWiFi, Bluetooth
Power SupplyAC power
Dimensions26.8 cm x 22.4 cm x 5.1 cm
PortsUSB, Ethernet
Supported ParametersSpO2, Pulse Rate
CompatibilityMasimo sensors
ModelRoot

Related product manuals