Root NIBPT Chapter 15: Specifications
www.masimo.com 142 Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an
environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV
contact
+8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+2 kV for
power supply
lines.
+1 kV for
input/output
lines.
---
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+1 kV -
differential
mode
+2 kV -
common mode
---
Mains power quality should be that
of a typical hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
100% for 0.5
cycle
60% for 5
cycles
30% for 25
cycles
100% for 5
seconds
---
Mains power quality should be that
of a typical commercial or hospital
environment. Root provides a
battery for continued operation
during power mains interruption for
a maximum of 4 hours.
Power frequency (50 /
60 Hz) magnetic field.
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of
typical location in a typical hospital
environment.