Root NIBPT Chapter 15: Specifications
www.masimo.com 136 Masimo
NIBP Pressurization Ranges
Weight Patient
Category
Initial
Pressurization
Maximum
Pressure
Greater than 75 lbs (34 kg) Adult 160 mmHg 280 mmHg
Between 15.4 - 75 lbs (7 - 34
kg)
Pediatric 140 mmHg 280 mmHg
Less than 15.4 lbs (7 kg) Neonatal 90 mmHg 140 mmHg
Electrical
Root
AC Power requirements 100-240 VAC~, 47-63 Hz, 180 VA (Max)
Fuses (2) 2 Amp, Time Delay High Breaking Capacity (5x20mm), 250V
Battery
Type Lithium Ion
Voltage 10.8V (Nominal)
Capacity 4 hours*
Maximum Charging Time 4 hours
*This represents approximate run time at the lowest brightness, using a fully charged battery.
Environmental
Root
Operating Temperature 50°F to 104°F (10°C to 40°C)
Transport/Storage Temperature -4°F to 122°F (-20°C to 50°C)
Operating Humidity 10% to 95%, non-condensing

Table of Contents

Other manuals for Masimo Root

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Root and is the answer not in the manual?

Masimo Root Specifications

General IconGeneral
CategoryMedical Equipment
Display10.1 inch touchscreen
ConnectivityWiFi, Bluetooth
Power SupplyAC power
Dimensions26.8 cm x 22.4 cm x 5.1 cm
PortsUSB, Ethernet
Supported ParametersSpO2, Pulse Rate
CompatibilityMasimo sensors
ModelRoot

Related product manuals