EasyManua.ls Logo

Masimo Root - Page 157

Masimo Root
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 155 Masimo
Procedure
1. Connect the manometer, 500mL bottle, and hand bulb to the NIBP module using
"T" adapters and connection tubing.
2. Power ON the manometer.
Set the unit of measurement to mmHg.
Verify the manometer has been calibrated.
3. Power ON the Root.
4. On the Root home screen, press the Main Menu icon .
5. Press NIBP .
6. Press CALIBRATION .
For the password, enter 4258.
Press the Enter key.
7. Press AIR LEAK TEST .
8. To begin the air leak test, press Test.
Wait for the countdown timer to reach 0 second.

Table of Contents

Other manuals for Masimo Root

Related product manuals