Page #225 background image
DICOM 12-13
Name Description
AE Title
Application Entity title, Here, it should be consistent with that of the
storage commitment server.
Port
DICOM communication port, 104 by default. Here, the port should be
consistent with that of the storage commitment server port.
Maximum Retries Reserved feature.
Interval Time(Sec) Reserved feature.
Timeout(Sec)
Value: 560s, in increments of 5s, and 15 by default.
Associated
Storage Service
The associated storage server is preset before storage commitment,
only after the exam is sent out, can storage commitment be created.
Add Add the DICOM service to the service list.
Cancel Click to cancel the parameter setting.
Update
Select an item in the service list, change the parameters in the above
area, and click [Update] to update the item in the service list.
Delete Click to delete the selected service in the service list
Default
Select an item in the service list, click [Default] and you can see “Y”
in the Default column.
Verify
Click to verify if the two DICOM application entities are normally
connected.
Exit Click to exit the screen.
12.1.4.6 Query/ Retrieve Setting
1. Enter the “Query/ Retrieve” page: “[Setup] [Network Preset] [DICOM Service]
Query/ Retrieve.
2. Select device, enter the information. For device setting, please refer to “12.1.3
DICOM Server Setting”.
z Click [Add] to add the service to the Service list.
z Select an item in the service list, change the parameters in the above area, and
click [Update] to update the item in the service list.
z Select an item in the service list, and click [Delete] to delete the service.
z Select an item in the service list, and click [Default] to set the server to be the
default service.
3. Select an item in the service list, and click [Verify] to verify the connection.
4. Click [Exit] to confirm the setting and exit the page.