100
9−3−6. Communication error process
The instrument returns the error command to the host side at the time of Communication error
or Execution error.
2468
@ 0 0 E R 0 1 Terminator
Error Command No.
Error
Command No.
Contents Remarks
01 Error of impossible
condition of execution
In the case of Function mode, Calibration mode,
Fine adjustment mode and Setting mode.
02 Another error caused
from the instrument
In the case of impossible to execute the receiving
command.
10 Parity error In the case of detecting the parity error.
11 Framing error In case of detecting error for stop bit .
12 Overrun error In the case of reading error for receiving
command.
13 Error of data code, data
length error
The receiving data code and data length are not
the same.
14 No applicable command The receiving command isn’t the same.
● In case that the Completion code (terminator) is not detected, the
error code will not be returned.
● In case that the communication error command is returned from the
instrument, consider its remedy at the host side.