7−13. Lock of key function
The instrument prepares the lock of key function.
With the Function F−06, execute the lock OFF of key function by setting each digit=0, and
executes lock ON of key funciton by setting each digit=1. As for default, “all is lock OFF of key
function”. Besides, the correspondence between the target of key funciton and setting digits are as
follows:
10
0
digit :ON/OFF of CHECK value(Press the key in the condition of pressing the
key.)
10
1
digit :ON/OFF of Peak Hold function, A/Z ON(), Reset of the peak value,
A/Z OFF(
)
10
2
digit :Reading out the change mode of S1 setting(), Reading out the change mode of
S2 setting(
)
10
3
digit :Execute the zero setting(Press the key in the condition of pressing the
key.)
10
4
digit :Reading out the function mode()
Moreover, when the reading out the function mode is locked, the function mode can be
proceeded after pressing the
key for over 2 seconds in the condition of pressing the
key for over 1 second.
7−14. CHECK value
By press of key together with the press of key in the Measurement mode, CHECK value
equivalent to the set value with the Function F−11 will be ON. And at the same time when the “
CHECK” on status display lights up, the load value equivalent to the set value with Function F−11
will be added to the load display value. As for default, 0.4 mV/V has set.
By press of the
key together with the press of key again, the CHECK value will be OFF
and returns to the former condition as it is.
● When the setting of Function F−11 is “00000”, load display value will
not vary even if the “CHECK” lights up on the status display.
(Because CHECK is 0.0 mV/V.)