35
524. Zero fine adjustment
(1) Procedures by key operation
During the Tare weight cancellation(A/Z) or Zero set is executed, and
also during effective for Zero tracking and Peak ON, Zero fine
adjustment mode can’t be entered into. (Displays ER5.)
The Zero fine adjustment can be entered after Tare weight
cancellation clear(A/Z OFF), cancellation of the Compensated data at
Zero set (Execution of F98), set OFF the Zero tacking (Set the
F08 and F09 to “00000”.) and set the Peak OFF.
During the calibration procedure, press the key to interrupt the
calibration. The calibration data will keep the same condition as it is
entered before, then returns to the Measurement mode.
When the key is pressed with the load display of “FUNC”, the
display will change as the following arrow marks indicate at every
time the key is pressed. However, every time the is pressed, the
display will change as the reverse direction of the following arrow
marks.
“FUNC”“CCAL”ACAL”“LCAL”“ZERO”“SPAN”
“TARE”“CHEK”“MONT”“VCAL”“VADJ”“FUNC”
“CCAL”→・・・・・(Hereinafter, it will repeat.)
Procedures
1
Press the key for about one second.
The load display will show “FUNC”.
2
Press the key four times.
It will make the load display proceeded as
“FUNC”“CCAL”ACAL”“LCAL”
“ZERO”.
Here, set the initial load application.