34
Procedures
Press the key.
The “SPAN” on the load display will flash on
and off, and span adjustment can be started.
Warning Take care not to apply load
variations due to vibrations and so on.
If load variation is applied, the span will
not stabilized, in due course there is a
possibility that correct reading of span
won’t be obtained.
12
When completed, the display on the load display
will show “END”.
However, when the value corresponding to the
maximum display value does not satisfy the
range from 0.4 mV/V to 3.1 mV/V, the right
Error code will light up for about 2 seconds, then
the display on the load display section will show
“SPAN”, and then returns to the step 10.
SPLSpan Over
SPHSpan Over
Error code
Error code
Press the key.
After “LCAL” mode is over, the load display will
h
th
t
ld
13
sho
w
th
e
present load.